INSTRUCTION MANUAL CODE 80260 REV A
Fig. 5.3
Fig. 5.4
D2
Lamp
1
Fig. 5.5
J
.P. SELECTA s.a.u.
6.3.2 Replace Lamps
High temperature!! Wait 20 minutes before open the lamp chamber after
power off to avoid scald!
•
Step 1, Tools Preparation:
Prepare a 6×150mm Cross Blade screwdriver and a pair of glove.
•
Step 2, Power Off:
Switch off the power supply and unplug the socket.
•
Step 3, Open the Cover:
Unscrew the 4 screws indicated in Fig. 5.3 (each side with 2 screws) and remove
the cover.
Light
chamber
•
Step 4, Open the Cover of the Light Chamber:
Unscrew the 2 screws on the light chamber cover and remove it (Fig. 5.4).
2
•
Step 5, Replace the D2 Lamp:
Unscrew the 2 screws on the D2 Flange (No.1 in Fig. 5.5), unplug the connector in
the power board (No. 2) and remove the D2 lamp. Draw on the cotton glove and
replace a new lamp. Fix the 2 screws and plug the connector again.
•
Step 6, Replace W Lamp:
Remember the direction of the filament before pull out the W lamp. Be sure
that the new lamp's filament is in the same direction as before.
Pull out the defected W lamp and draw on the cotton glove. Insert the new W lamp
as deep as possible on the lamp seat. Be sure to keep the filament in the same di-
rection as the old one face.
W
Lamp
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
Feb/2017
(It can be modified without notice)
Fig. 5.6
Page: 54
Filament
Tel 34 93 770 08 77 Fax 34 93 770 23 62