press F3 to go into «Process» screen. Input the values of «Begin Time», «End Time»
and «Factor» separately and press [ENTER] to confirm. Then the value of «I.U.»
would be displayed on the screen (Fig.4.14).
•
Step 8, Search:
After scan finished, press F4 to go into search mode. Press
value of every point (Fig. 4.15).
•
Step 9, Print:
Press [PRINT] to print curve.
•
Step 10, Delete the Curve:
Press [CLEAR] to delete the local curve.
•
Step 11, Save the Curve:
After scan, press [SAVE] to save, input the file name by the numeric keypad. Press
[ENTER] to confirm.
5.3.5 DNA/Protein Mode
•
Step 1, Go into DNA/Protein Test:
In the Main menu, press [5] or
press [ENTER] to confirm (Fig. 4.16).
•
Step 2, Setup Coefficients:
Press F1 to set coefficients, input all the values of F1 to F4 by means of the numeric
keypad according to the indication and press [ENTER] to confirm.
•
Step 3, Choose Measure Method:
Press F2 to set method. Press
or «Absorbance Difference 2» followed by [ENTER] pressed to confirm. If users
don't want to measure reference, use
[ENTER] pressed to confirm the choice.
•
Step 4, Set Concentration Unit:
Press F3 to set concentration unit. Use
[ENTER] to confirm.
•
Step 5, Calibrate 100%T/0Abs:
Place the Reference Sample in the light path, press [100%T/0 ABS] to calibrate
100%T/0Abs.
•
Step 6, Sample Measure:
Place the sample to be tested in the light path, press [START/STOP] to measure.
The result will be displayed in the data sheet (Fig. 4-17).