WARNING
Always allow the product to reach room temperature before you begin setup or test functional operations, to
prevent permanent product damage.
Do not use this product if it has identifiable failures, defects, malfunctions, or damage.
Do not use this product under any conditions if use would cause injury to the operator or patient.
Only operate the product when all operators are clear of the mechanisms.
To avoid the risk of electric shock, this equipment must be connected to a supply mains with protective earth.
Always store the power cord to avoid the risk of entanglement, damage to the power cord, or potential shock
hazards. If the power cord is damaged, remove the product from service and contact the appropriate maintenance
personnel.
Always allow enough clearance between the head end of the product and the adjacent wall, so you can unplug the
power cord from the wall outlet during an emergency.
Always store the power cord before transporting the product.
Do not attach the power cord to any parts of the product.
Always unplug the power cord and call maintenance if unanticipated motion occurs.
Do not store items under the bed.
Do not use the bed without the litter covers.
CAUTION
Do not place objects into any gaps of the product.
Note: Product has suitable mains to isolate its circuits electrically from the supply mains on all poles at the same time.
Before placing the product into service, make sure that these components work:
1.
Visually inspect the product for any signs of shipping damage.
2.
Make sure that the product and all components and accessories have arrived.
3.
Depress the brake pedal and make sure that the brake, steer, and neutral positions operate.
4.
Raise and lower the siderails to make sure that they move, stow, and lock securely in the full up position.
5.
Plug the battery cable into the control box
6.
Plug the power cord into a protective earthed outlet
7.
Press each button on the siderail control panel, nurse control pendant, and patient control pendant (option) to make
sure that each function operates
8.
Make sure that the battery is fully charged.
9.
Make sure that the CPR release handle operates.
10. Make sure that optional accessories are installed and operate as described in this manual.
www.stryker.com

Setup

(Plugging or unplugging the battery cable on page
(Plugging or unplugging the product on page
(Nurse control pendant on page
KK-7500 REV 02
1-16).
1-31).
English
EN