08
THE ORDINARY MAINTENANCE CONSISTS FIRST OF ALL IN CLEANING:
Cleaning must be careful and done after every use
Important:
Cleaning the accessories and the device allows a correct working of the machine and the
protection of both the patient's and the operator's health
INSTRUCTION FOR CLEANING AND DISINFECTION OF THE UNIT:
Wear latex gloves
-
External:
Remove all the accessories and clean the aspirator with a wet cloth with a not aggressive detergent
-
Pots:
-
Empty the pot out (the pot's cleaning can be facilitated by putting some water inside: the aspired
diluted liquid will be more easily eliminated)
-
Disassemble every part,(the valve "too full" must be correctly reassembled, the float must have the
gasket turned upward and the wrapping must be well inserted)
Tubes:
Wash by running water
-
Check that tubes don't present dirty parts with presence of corpuscles
-
The tubes can be steam sterilized at 134°c, cold disinfected, or following the sterilization protocol of
-
the site where the unit is used
Filter replacement:
-
For a correct working, check the filter and replace it when necessary, in order to avoid every inner
pollution which would make it unusable.
-
The filter cannot be cleaned, it must be replaced when it is choked
The accessories (pot, filter...) must be supplied by Tecno-gaz
-
Temperature of stockage of the filter +5°c/ +40°C
-
THE EXTRAORDINARY MAINTENANCE IS NOT EXPECTED. IN CASE OF ANOMALIES CONTACT
TECNO-GAZ FOR TECHNICAL ASSISTANCE AND REPARATION
WARNINGS FOR SHIPMENT:
The unit must be SANITIZED with particular attention to parts retuned for assistance.
The package must be suitable, the best think is preserve the original package.
DB1F020_rev8.docx
ORDINARY AND PROGRAMMED MAINTENANCE
24