Analysis And Research Of Failures; Commercial Data, Spare Parts And Documentation - Tecno-gaz TECNO40 Istrucciones Para El Uso

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

09
N° FAULT
CAUSE
IT DOESN'T
1
SUCK
THE FLOAT
2
DOESN'T
3
NOT WORKING
10
The positions refer to the electro-pneumatic outline (following chapter).
DESCRIPTION
POWER SUPPLY CABLE
KIT OF PIPES FOR EXTERNAL WIRING
HYDROFOBIC FILTER
POT WITH NON RETURN VALVE 2 L
POT WITH NON RETURN VALVE 4 L
DB1F020_rev8.docx

ANALYSIS AND RESEARCH OF FAILURES

THE
POT
IS
BADLY
SCREWED DOWN
EXTERNAL
PIPES
FOLDING
FILTER IS OBSTRUCTED
WRONG ASSEMBLING
EXTERNAL FUSE
INTERRUPTED

COMMERCIAL DATA, SPARE PARTS AND DOCUMENTATION

ASSISTANCE
UNSCREW AND SCREW THE TOP, CHECK
GASKET'S INTEGRITY AND RIGHT POSITION
WITH
REMOVE NARROWING ON THE PIPES
FILTER MUST BE REPLACED
CHECK THE CORRECT POSITION OF NON
RETURN VALVE
IN CASE OF NOT WORKING, ADDRESS TO
TECNO-GAZ
REFERENCE
CE06006
SB1A138
CPF0029
CM84151
CM84152
25