8.2
Wireless footboard (on request)
The Wireless footboard (on request) has a rechargeable battery.
The LED (Pos _7_ fig.1) starts flashing when the battery is low; to recharge the battery,
simply connect the battery charger provided with the footboard and press any button; the
LED starts flashing slowly and remains steady when the battery is fully charged; the time
for recharging depends on how low the battery is.
N.B. It is advisable to recharge the footboard at the end of every working day.
Make sure to connect the battery the right way round
9

MAINTENANCE

CLEANING THE CHAIR.
The equipment is made from materials that can be cleaned and disinfected as indicated in
these instructions.
The many types of pharmaceuticals and chemical products used in the surgery can
damage the painted surfaces and the parts made from plastic.
The tests and research carried out revealed that the surfaces cannot be fully protected
against the aggression of all the products available on the market.
The effects of these products essentially depend on the length of time these remain on the
surfaces.
It is important to remove these substances immediately from the surfaces of the unit, using
a damp cloth.
If using detergents, spray these onto a cloth and not directly on the unit.
ATTENTION!
Do not use products containing alcohol, ammonia, abrasive substances or benzol. Do not
exceed the maximum allowable concentrations shown below.
Quaternary ammonium: maximum 2%Glutaraldehyde: maximum 2%
TECNO-GAZ CANNOT BE HELD LIABLE FOR DAMAGE DUE TO THE USE OF
DISINFECTANTS CONTAINING ACTIVE SUBSTANCES OTHER THAN THOSE LISTED
ABOVE.