7. The cartridge door will not
close.
8. The probe will not sit
correctly in the chamber.
9. The cycle will not start.
10. Liquid is leaking from the
trophon2.
11. The trophon2 is failing
multiple cycles.
12. The trophon2 takes too long
to warm up between cycles
L03031 1.0 03/2018
Correct cartridge type has been inserted.
Cartridge is correctly positioned.
Cartridge lid has been removed.
The probe is validated for use in the trophon2 – see section
A1.2.
The probe is loaded correctly.
The IPP is correctly engaged – see section C3.3.
The probe is loaded correctly.
The chamber door is closed.
Confirm probe is clean & dry before pressing start.
All on screen instructions have been followed correctly.
WARNING: Any fluid leaking from the trophon2 may
contain hydrogen peroxide.
If liquid or mist is seen coming from the trophon2 at any time:
1. Do not come into contact with the mist or liquid.
2. Wear appropriate PPE.
3. Ensure area is well ventilated.
4. Allow the trophon2 to complete the cycle.
5. Turn off the trophon2 and remove the power cord.
6. Contact your customer service representative.
7. Consult the Safety Data Sheet.
Record any error codes and colour of the Chemical Indicator
and contact your customer service representative.
Ensure that the probe is removed promptly at the end of each
cycle.