Menu
Records then choose to view one of the record list options detailed below and follow the on
screen instructions.
Last Cycle*
View on screen or print the details of the most recent HLD cycle.
The below logs must be downloaded to USB for viewing
Disinfection Cycle Log
Past disinfection cycles.
NOTE: The Operator ID listed (if AcuTrace enabled) is the user responsible for logging the pass or fail
Chemical Indicator result via the Operator Card.
Disinfectant Log**
Disinfection cartridges scanned and used by the trophon2.
Chemical Indicator Log**
Chemical Indicator packs scanned and used by the trophon2.
USB Export
The records above can be downloaded onto USB. Add a USB drive into any of the three ports on the
left hand side of the trophon2 and select the required record, then select Export and follow the on
screen instructions.
*Not available for USB export
**AcuTrace must be enabled to access this record
PART E –MAINTENANCE AND ROUTINE CARE
SECTION E1: Preventative Maintenance Service
Your trophon2 requires annual preventative maintenance and a major service at every 5,000 cycles.
Once the service due reminder screen message is displayed, please contact your customer service
representative to arrange a service. Contact details, if not provided by your distributor can be found on
the Nanosonics website.
The service options found: Menu
should only be used when advised by service personnel.
SECTION E2: Purge Cycle
The purge cycle removes and converts any remaining disinfectant from the trophon2, into oxygen and
water.
E2.1 When to Run a Purge Cycle
Prior to transporting the trophon2, if it has previously been in use.
When an on screen message states that your trophon2 requires purging. This will occur upon
disinfectant expiry at 30 days after insertion. Follow the on screen instructions to purge.
NOTE: Purging can be deferred until convenient on the day of expiry.
Initiate a purge cycle:
When the trophon2 detects an error that requires a service call.
Before lifting or moving the trophon2.
For troubleshooting purposes when directed by service support only.
E2.2 How to Initiate a Purge Cycle
NOTE: Once the purge cycle has started it may be paused but it cannot be cancelled. Do not switch
the trophon2 off during purging as this will restart the purge cycle. Do NOT attempt to open the chamber
or cartridge door during the purge cycle.
To initiate a purge cycle:
L03031 1.0 03/2018
Maintenance are used to troubleshoot device malfunctioning and
19