6.3.2. ELECTROMAGNETIC COMPATIBILITY AND SUSCEPTIBILITY
Compliant with IEC 61326-1 for emissions and immunity in industrial use.
6.3.3. PROTECTION FROM ELECTRIC SHOCKS
Type A clamp with double insulation or reinforced insulation in the grip part in normal use, and double
insulation or reinforced insulation between the primary and the secondary output and the terminations
part of the USB power supply and the BNC output.
Maximum service voltage per EN 61010-2-032
- 300 V in installations of category III and degree of pollution 2.
- 600 V in installations of category II and degree of pollution 2.
6.3.4. VARIATIONS IN THE NOMINAL OF THE USE
Qauntity of influence
USB supply voltage
Relative humidity
Temperature on measurement
Setting time
Pass band
(without filter)
Position of the cable
Repeatability of the
measurement
Common mode rejection
Rejection of an external current
at 50 Hz
1
External magnetic field
(spatial position)
2
High-frequency
electromagnetic field,
10 V/m
between 650 MHz and 750 MHz
(1) Conductor outside the jaws and parallel to the position of a conductor in the jaws.
(2) Measurement made after demagnetization of the clamp. For this purpose, set to AC, I = 0, gradually
increase the value to I = 40 A. Gradually reduce the value to I = 0.
Influence
5 VDC ± 5 %
or storage battery voltage
between 20 % and 100 %
charge
10 % at 90 % RH
-10 °C at +45°C
45°C at +50°C
upon start of operation
DC to 1 kHz
to 100 kHz
to 500 kHz
to 1 MHz
DC to 100 Hz
10 open./clos.
50 Hz
400 Hz
near air gap
jaws
grip
DC (Earth's)
value of the current
28
Typical
0
± 0.12 %/°C
± 0.22 %/°C
10 s
-0.2 dB
-0.5 dB
-1.5 dB
mean Δ: 0.3 %
> 90 dB or 0.003 %
> 80 dN or 0.01 %
30 dB
34 dB
40 dB
± 0.3 mV or
± 30 mA
1.2 A
Maximum
none
0.1 % of Is
1 %
-0.5 dB
- 1 dB
-3 dB
± 1.5 %
± 1 mV or
± 100 mA
1.5 A max.