Variations in the nominal working range.
Distortion quantities
Operating
position
Temperature
Humidity
Supply voltage
Frequency (on voltmeter)
Electromagnetic field
(1)
For measurements in MΩ, it is a % of the measurement.
For the other measurements, it is end of scale.
(2)
For insulation measurements from 0.1 to 100 MΩ, voltage measurements from 0 to
600 V, continuity measurements from 0 to ± 10 Ω and resistance measurements from
0 to 1000 Ω.
(3)
The magnetic fields envisaged are fields with stable amplitude and direction, and of
frequency between 0 and 60Hz.
ELECTRICAL SPECIFICATIONS
Power supply
The instrument is powered by 4 batteries 1.5V type R6 alkaline
The voltage range ensuring correct operation is from 4.5V to 6.5V
Check correct operation by testing the battery (see "Use").
Consumption is approximately :
300mA for R = 0.5 MΩ on the MΩ 500V range
600mA for R = 1 MΩ on the MΩ 1000V range (C.A 6513 only)
200mA on continuity on the +10 Ω and -10 Ω ranges
10mA on ohmmeter on the 1kΩ range (C.A 6513 only)
The average service life is :
1000 insulation measurements of 10s on the MΩ 500 V range for R = 500 kΩ,
200 insulation measurements of 10s on the MΩ 1000 V range for R = 1 MΩ (C.A 6513 only)
1500 continuity measurements of 10s on the 10 range.
Limits of working
range
0 + 90 °
0 - 90 °
-10 to + 55 °C
3 % / 10°C (1)
20 to 80 % RH (2)
the measurement
4.5 to 6.5 V
the measurement
45 to 400 Hz
0 to 400 A/m (3)
19
% variation of
the measurement
Typical
Max.
-
≤ 5 % of
-
the measurement
5 % / 10°C (1)
5 % of
10 % of
the measurement
0.1 % of
0.2 % of
the measurement
-
≤ 0.1 % of
the measurement
-
1/2 class