6.5. Variations in nominal field of use
Influence
quantities
Temperature
Relative
humidity
Supply
voltage
Frequency
(inV and in mA)
Capacity
en parallel
on the resistance
6.6. Limits
The instrument is protected across all ranges against a constantly
applied voltage, between any two terminals, of 600 V AC/DC on
the 1039, and 720 V AC/DC on the 1035.
The 1039 is protected for 10 s against an accidental
overvoltage of 720 V across all the ranges.
The current input on the 1039 accepts 0.63 A, beyond which
it is protected by fuse.
6.7. Construction specifications
Overall dimensions of the unit (L x l x h): 211 x 108 x 60 mm
Dimensions of the display: 73 mm x 54.3 mm
Weight: approx. 835 g
Materials:
- Polycarbonate housing
- Crystal polycarbonate screen
- Elastomer external mouldings
- Silicon keyboard.
Stand:
Enables the instrument to be tilted at 30°. It clips onto the
bottom of the housing when not in use.
6.8. Compliance with international standards
Electrical safety: IEC 1010-1 + A2 (Nov. 95),
Dual insulation:
Pollution level: 2
Installation category: III
Rated voltage: 600 V
Operating
range
limits
-10 to + 55°C
20 to 80% HR
7 to 10 V
DC and
15..400 Hz
0 to 5 µF at
nominal current
IEC 61557 (Feb. 97) and DIN EN 61557.
Measurement
variations
Typical
(1% R ±1 ct)/10°C (2% R ±2 ct )/ 10°C
2% R ± 2 ct
(1% R ±1 ct)/V
1% R ±1 ct
Negligible
40
Maximum
3% R ± 2 ct
(2% R ±2 ct)/V
2% R ±2 ct
1% R ±1 ct