11. MAINTENANCE
CAUTION: DISCONNECT ALWAYS THE MAIN SUPPLY BEFORE EACH TECHNICAL
INTERVENTION!
DAILY ROUTINE MAINTENANCE
11.1.
Clean the seal on the door
the gasket using a wet cloth or the soft part of the supplied sponge.
This cleaning operation must be carried out to remove possible dirt which can cause a
pressure loss in the sterilisation chamber and a possible seal cut.
Check the water level in the tank (see paragraph 8).
WEEKLY ROUTINE MAINTENANCE
11.2.
To remove the deposits on the chamber floor, clean with the abrasive part of the supplied
sponge after removing the tray-holder.
To moisten the sponge, only use distilled or demineralized water.
Empty the water tank (see paragraph.9).

QUARTERLY MAINTENANCE

11.3.
Lubricate the hinges and the closing pin with silicon oil.
Adjusting the door
If the handle
Photo F to increase the pressure of the gasket on the chamber.
and on the outer edge of the sterilisation chamber sealing
1A
:
1A
is too frail while closing and blocking the door, turn it counterclockwise
2A