4. SUPPLIED ACCESSORIES

The autoclave is sold with the accessories included in the article and contained in the same package.
Accessories provided:
N° PIECE
1
Tray extraction and door adjusting key
4
Tray
1
Instruments holder
1
Funnel
1
Tray holder
THE EQUIPMENT LISTED ABOVE DO NOT REQUIRE ADVICES
WARNING: THE COVER OF THE TANK IS IN ACCESSORIES KIT
5. INSTALLATION
The installation is a fundamental operation for the following use and good operation of the device.
Here is a list of the points to be followed in order to achieve a good installation.
1. The device must be installed inside a laboratory accessible only to the authorised personnel.
2. The room in which Autoclave is installed must be lighted according to standard UNI 10380 and
enough ventilated.
3. Position the device on a plane surface with minimum capacity of 60 kg.
4. The autoclave feet are adjusted by the producer, in order to obtain a correct level on plane.
Their modifications could compromise the autoclave good operation Photo A.
5. Leave at least 5 cm between the device rear part and the wall.
6. Position the autoclave at such a height as to make it possible for the operator to check the
whole sterilisation chamber and to carry out the normal cleaning operations.
7. Do not install the autoclave near washing basins, taps, because the device furniture is not
waterproof.
8. Do not place trays, papers, containers with liquids, etc. on the device, because the
device grids are used to ventilate and must not be obstructed.
9. Do not lean on the door
10. Do not install the device near heat sources (for instance: Autoclave, Bunsen burner, kiln); keep
a minimum distance of 40 cm.
11. In case the filling tank is directly discharged in waste waters, position the device at a height
which is higher than the discharge.
12. The device cannot be built in because the water filling in the tank is carried out from above.
13. Install the device in such as way as not to excessively bend the supplying cable.
14. Make sure that the electrical system to which the device is connected is produced according to
the laws in force and has the suitable dimensions for the device features; see PAR.2 of this
manual.
when it is opened.