TABLE OF QUALITY LEVEL IN ACCORDING TO DIRECTIVE UNI EN 13060 : 2009
Evaporation residue
Silicium oxide, SiO2
Iron
Cadmium
Lead
Rest of heavy metals except iron, cadmium, lead
Chloride (Cl')
Phosphate (P20s)
Conductivity (at 20° C)
PH value (degree of acidity)
Appearance
Hardness (E Ions of alkalin earth)
9. EMPTYING THE WATER TANK
WARNING: SHUT DOWN THE AUTOCLAVE WITH THE
DISCHARGING THE TANK!
Connect the supplied pipe to the discharge tap 3A.
1.
Unscrew the ring nut in counterclockwise direction Photo C (make 2 complete laps).
2.
Wait that all the tank is emptied.
3.
Rescrew the ring nut in clockwise direction Photo D.
4.
In case of sterilisation failure, water in the tank could contain contaminated rests; used
5.
protective lattice gloves to carry out this operation.
Photo C
CEN STANDARD UNI EN 13060 : 2009
10
≤
1
≤
0.2
≤
0.005
≤
0.05
≤
0.1
≤
2
≤
0.5
≤
15
≤
5 to 7,5
Colourless clean without sediment
0.02
≤
MAIN BUTTON
BEFORE
Photo D
mg/l
mg/l
mg/l
mg/l
mg/l
mg/l
mg/l
mg/l
µs/cm
mmol/l