9 Apéndices
G.15 Linearity
Table D.3 in section D.3 of the IRMA System User Manual is replaced by the following in Table G.8:
Analyte
pH
pCO
mmHg
2
pO
mmHg
2
Sodium (Na + ) mmol/L
Potassium (K + ) mmol/L
Ionized Calcium (iCa) mmol/L
Hematocrit (Hct) %
G.16 Detailed Instructions
The dimensions of the videos displayed in the IRMA System user interface have been increased. The numbered instructions
are now scrolled beneath the video. To view the instructions as a list press the user manual icon.
Slope
Intercept
0.995
0.038
0.997
0.141
1.056
-11.72
1.010
-1.894
1.017
-0.065
0.994
-0.008
1.033
-0.489
G.10
Table G.8
R
Reference Range
2
0.999
6.514 - 8.131
0.998
0.75 - 230.2
0.994
6.15 - 778.2
0.996
89.5 - 218.7
0.999
1.68 - 10.62
0.999
0.18 - 6.42
1.031
7.8 - 79.3
Tabla de contenido