ENGLISH
06
Remove the packaging and check the conditions of the equipment.
Do not use equipment that presents evident damage due to transportation.
6.1
Installation of unit-mounted MASTER FLUX PLUS
Screw the 3 steel spacers (3) onto the flowmeter box (2).
Remove the protections from the flowmeter box supply nipples.
Install the flowmeter box on the unit by inserting the steel spacers (3) into the relative holes
on the unit, securing them to the unit using 3 plastic spacers (5), 3 washers (6) and the three
thumb nuts (7) supplied.
Position the plastic cover near the three holes on the unit.
Attach the plastic cover by screwing on the three thumb screws (4) supplied.
EXHAUST NOZZLE INSTALLATION (Figura 2 Pos 9)
If you are using the exhaled gas exhaust tube, screw the supplied nozzles onto the plastic
fittings.
CYLINDER INSTALLATION (NOT SUPPLIED) (Figure 10)
Attach the reducers to the cylinders (see instruction manual for reducers).
For 5 lit cylinders install the shelf and attach it using the relative shelf stops.
For 10 and 14 lit cylinders position the shelf on the bottom of the unit.
The reducer pressure gauges (M) must be placed in such a way that they are visible to the
operator.
Lock the cylinders to the piece of furniture using the special Velcro strap supplied.
Secure the cylinders to the unit using the velcro strap supplied.
Screw the O2 reducer (R1) tube end nut onto the flowmeter box supply nipple (I) (RIGHT
THREADING).
Screw the N
O reducer (R2) tube end nut onto the flowmeter box supply nipple (H) (LEFT
2
THREADING).
Close the rear door with the two magnets located on the top of it.
14
INSTALLATION
FLOWMETER BOX INSTALLATION (Figure 2)
see enclosed Quick Guide 0ZMFI0007
FRONT COVER INSTALLATION (Figura 2 Pos 1)
see enclosed Quick Guide 0ZMFI0007