Electromagnetic compatibility
Electromagnetic emissions
Guidance and manufacturer's declaration – electromagnetic emissions for all EQUIPMENT and SYSTEMS
The Blood pressure monitor (BM 54) is intended for use in the electromagnetic environment specified below. The customer of the
user of the Blood pressure monitor (BM 54) should assure that it is used in such and environment.
Emission test
Compliance
RF emissions
Group 1
CISPR 11
RF emission
Class B
CISPR 11
Electromagnetic immunity
Guidance and manufacturer's declaration – electromagnetic immunity – for all EQUIPMENT and SYSTEMS
The Blood pressure monitor (BM 54) is intended for use in the electromagnetic environment specified below. The customer of the
user of the Blood pressure monitor (BM 54) should assure that it is used in such and environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Power frequency
(50Hz) magnetic field
IEC 61000-4-8
Electromagnetic environment – guidance
The Blood pressure monitor (BM 54) uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
The Blood pressure monitor (BM 54) is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
IEC
60601
Compliance
test level
level
±6 kV contact
±6 kV contact
±8 kV air
±8 kV air
3A/m
3A/m
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic tile. If floor are covered
with synthetic material, the relative humidity should be at least 30%.
Power frequency magnetic fields should be at levels characteristic
of a typical location in a typical commercial or hospital environment.