Recommended separation distances between portable and mobile RF communications equipment and the
The electronic ENDOFLATOR
®
model 26 4305 20-1 is intended for use in an electromagnetic environment in which the RF disturbances are
controlled. The user of the device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the device, according to the output power of the communications equipment as recommended below.
150 kHz to 80 MHz
Nominal power of
D
transmitter
[W]
0.01
0.1
1
10
100
For transmitters whose maximum nominal power is not listed in the above table, the recommended separation distance d in meters (m) can be determined
using the equation belonging to the relevant column, where P is the maximum nominal power of the transmitter in Watts (W) as speci ed by the manufacturer
of the transmitter.
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. The propagation of electromagnetic quantities is affected by absorptions and re ections of buildings,