• The device must always be protected against wetness and mois-
• Surge category II
• Soiling degree II
Requirements of cable for mains voltage and sys-
tem components and installed lines
Use only cables and installed lines which meet the following require-
ments:
• Dielectric strength 30 V ... 260 V according to the nominal voltage
• The cable ducts installed by Gebr. Heyl in the device have a
• The terminal strips on the circuit board require core cross-sections
Optimum cross-section of the cores
NOTE
If the core cross-section is less than 0.5 mm
when the cores are loosened from the terminal strip. We therefore
recommend using wires with cross-sections greater than 0.5 mm

Requirements for cable ducts

• The recesses in the housing are intended for M16 ducts.
• The ducts should have a smooth and rounded opening (to protect
• Be careful to ensure securely fastened bending protection, which
• The duct should include a strain relief that prevents slippage of the
• You can order cable ducts from us as spare parts
ture. Under no circumstances may it be exposed to splashed water
or condensate.
(see specification plate)
clamping range of 4.5 mm – 10 mm. This means that the external
diameter of the laid cable must remain within the range of 4.5 mm
– 10 mm. If you use other ducts, the cable diameters must corre-
spond to the ducts.
2
of between 0.08 mm
and 2.5 mm
fine-wire cores with wire end ferrules without plastic collar.
For fine-wire cores with wire end ferrules and plastic collar, the
cross-section may be up to 1.5 mm
For single-wire cores, AWG28 – AWG12 can also be used.
against bending and abrasion).
must be five times the length of the maximum cable diameter.
cable and that cannot be disengaged without a tool.
accessories).
If you use another cable duct, the material must have a flammabil-
ity rating of V1 or better.