Setting Indicator Fill Level; Cleaning Of The Measuring Chamber And The Viewing Windows; Cleaning The Filter Housing - Testomat EVO TH Serie Manual De Instrucciones

Dispositivo automático de análisis en línea para dureza total residual (dureza del agua)
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Servicing and maintenance
REAGENT
Filling level
!
ATTENTION
70

Setting indicator fill level

The indicator fill level must be set when refilling the indicator bottle
set or reset during a replacement.
Select the menu
=> Service => Reagent => Filling level
100%
100%
With "OK", the current fill level is reset to 100 percent.
If only the fill level should be reset, confirm the fill level with "OK
or.
Change the value of the fill level with the cursor keys and subse-
quently confirm with "OK".
Cleaning of the measuring chamber and the view-
ing windows
Cleaning interval: ¼ yearly
Switch off the device or press the "Pause" button
Check that the measuring chamber is completely emptied.
Connect the manual valve of the secondary line to the Testomat
EVO TH.
Release the spring lock
and take it out.
Remove both the viewing window holders
viewing windows for cleaning.
Remove the film on the viewing windows with isopropanol.
Clean the measuring chamber with 10 percent hydrochloric acid
and then rinse it thoroughly.
Reinsert the viewing windows after cleaning and fix them in place
with the viewing window holders
Do not forget to insert the O-ring seals and be careful to ensure
correct fitting in the groove.
Put the measuring chamber back in place by tilting the unit and
secure it with the spring lock.
Removal and installation of the sight glass windows
Ensure tension-free mounting of the sight glass windows. Tighten the
screws equally alternating both sides. Otherwise, the sight glass win-
dows may break.

Cleaning the filter housing

Cleaning interval: ½ yearly
Connect the manual valve of the secondary line to the Testomat
EVO TH.
Press the "Pause" button and wait until the current measurement
has finished.
, tilt the measuring chamber upwards
and then remove the
®
®
Tabla de contenido
loading

Este manual también es adecuado para:

Evo th 2005Evo th 2025Evo th 2100Evo th 2250

Tabla de contenido