Input basic programming data
NOTE
Timing control
Internal trigger by timer.
MODE OF OPERATION
Time controlled
Volume interval
Volume and Time
Quantity control
Triggered by water meter
MODE OF OPERATION
Time controlled
Volume interval
Volume and Time
INTERVAL
Time
Volume
03501
34
Input basic programming data
Delayed reaction
During analysis, the response to any key presses may be delayed.
Select operating mode
Under the menu option
type of analytical control. The Testomat
control timing or quantity via water meter or a combination of both.
Smallest pause interval = 0 minutes between analyses. Longest
pause = 99 minutes.
The analysis interval (interval between two analyses) is determined
by the configured purge cycle, the programmed pause time (interval)
and the duration of analysis. The duration of analysis is directly de-
pendent on the measurement value.
Select timing control
Select in the menu =>
operation
*
Confirm the selection with "OK"
(An " * " appears at the end of the line)
(The factory default is "
Now enter the interval.
Select quantity control
Smallest interval = 1 liter, largest interval = 9999 liters. The analysis
is performed once the programmed quantity of water has passed.
Before the analysis, the line and measuring chamber are purged
(observe programmed purge cycles).
Select in the menu =>
ation
Confirm the selection with "OK"
(An " * " appears at the end of the line)
Select in the menu =>
*
Volume
Confirm the selection with "OK"
Input the corresponding flow volume in liters
10m
Confirm the input with "OK"
Now select the
"Operating mode"
Basic program
Time controlled
=>
Time controlled
Basic program
Volume and Time
=>
Basic program
water meter
.
you can select the
®
EVO TH gives you scope to
Mode of
=>
" " * ")
Mode of oper-
=>
Interval
=>
=>