Connect Inputs And Outputs - Testomat EVO TH Serie Manual De Instrucciones

Dispositivo automático de análisis en línea para dureza total residual (dureza del agua)
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

!
CAUTION
Terminal
No.
designation
13
Out +
14
Out -
Ext. Ack.
15
(external
16
acknowledge
)
17
WM
18
(watermeter)
19
Stop
20

Connect inputs and outputs

®
The Testomat
EVO TH device has the following sockets for control
and monitoring functions. Proceed as follows for connection:
Loosen the strain relief of the cable ducting (union nut).
Take out the blanking plugs of the corresponding cable ductings.
Push through the cable of the component.
Tighten the union nut of the cable ducting and so establish the
strain relief.
Re-secure the upper door after installation with both fixing screws.
Correct connection of the inputs and outputs
Do not expose the connections to any external voltage!
Ensure that the cores in the terminals are securely in place
Incorrect connections will damage the device!
Typ
Function
e
OUT Current loop 0/4 - 20 mA
External reset / acknowledgment input
IN
port for error and alarms
Water meter intake
IN
Common ground for inputs
External disruption of analysis
IN
Common ground for inputs
A detailed description is included in
puts/outputs".
Note
Galvanically isolated
Break contact/closing
contact programmable;
only connect isolated
break contact/closing
contact
Only connect isolated
break contact/closing
contact or observe the
technical data of the me-
ter!
Only connect isolated
break contacts/closing
contacts!
"Description of signal in-
Assembly
23