CAUTION
Connect indicator bottle
Flawless operation of the Testomat
guaranteed when Heyl Testomat 2000
use of external indicators may invalidate the guarantee.
Insert indicator bottle
Open the lower housing door by pulling on the right side
Remove the closing cap of the indicator bottle
Remove the plastic bag from the inner side of the lower housing
door. It contains the screw cap with hole
the screw cap
Join the parts together as shown below
Screw the hose connector
the insert
Plug the insert with screwed-in suction hose into the indicator bot-
tle
Now screw the screw cap with hole
bottle
Open water intake
Open the lower housing lid
Slowly open the manual stop valve to prevent the measuring
chamber from overflowing. Configuring the flow regulator will take
some time during the initial commissioning.
Check the leakproof quality of water-conveying parts
If any water sprays out of the hose
ber
, restrict the water intake to some extent with the manual
stop valve. The measuring chamber should be filled within two to
six seconds!
Automatic venting
After the device is switched on the indicator lines are automatically
vented and the measuring chamber purged, until indicator is detected
in the measuring chamber. The automatic venting cannot be inter-
rupted. Wait until the venting process has ended and acknowledge
the "voltage failure" error message by pressing the horn key. The
device is then ready for use.
®
EVO TH device is only
®
indicators are used! The
and the insert
of the suction hose
hand-tight onto the indicator
of the measuring cham-
Assembly
for
hand-tight into
25