Testomat EVO TH Serie Manual De Instrucciones página 63

Dispositivo automático de análisis en línea para dureza total residual (dureza del agua)
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Error message
36 Fault analysis
37 Indicator low
38 Water low
1 2
39 Ext. light influence
40 BOB not possible
66 Fault auto remove
air
Description, possible
causes
1
No correct analysis, e.g.:
Air in dosing hoses?
Insufficient mixing
Indicator expiry date
exceeded
Foreign indicator in the
device
Indicator minimum quan-
tity is not attained
- without BOB: 10%
- with BOB: according to
calculation
No water intake despite
permanent lamp "IN"
Intake pressure too low
Water remains in the
measuring chamber, de-
spite the lamp "OUT"
lighting up
Photo sensor not dark-
ened
Permanently illuminated
LED in measuring
chamber
The quantity of indicator
no longer suffices for 72
hours operation without
supervision
Automatic venting has
failed.
1
Device reverts to pause mode until the error is acknowledged
2
The error
Water shortage
is the only case in which the number of consecutively occurring errors
in the menu =
be adjusted until the relevant alarm or message is triggered. One
error is generated per measurement cycle.
Solution, troubleshooting measures
Re-tighten dosing pump connections
Renew intake insert in bottle
Check suction and pressure hose for dam-
age
Replace stirring bar
Replace indicator, only use Heyl Testomat
2000
Check indicator fill level,
bottle
Check water inflow
Plug at the inlet valve corroded
Clean filter sieve
Replace valve block
Remove the pressure regulator core
Outlet valve blocked or defective. Clean or
replace valve
Shut the device door
Avoid sunlight
Hardware problem. Send device for repair.
Insert
Possible causes can be problems with the
dosing pump, optics, cloudiness, dirt contam-
ination, analysis error or water shortage. To
determine the error more accurately, keep
the OK button pressed and restart the de-
vice. Another error message is displayed.
Proceed to troubleshooting, as described in
the error message.
is treated as a special case here, as this
>Basic program => Water low count
Error messages/troubleshooting
®
indicator
Insert new indicator
and
Reset indicator fill
new indicator
level.
can
63