Ext. Ack.
Terminals 15,16
WM
Terminals 17,18
WATER METER
1 l/Impulse
2,5 l/Impulse
5 l/Impulse
10 l/Impulse
100 l/Impulse *
500 l/Impulse
1000 l/Impulse
OUT
Terminals 13,14
NOTE
External delete (external acknowledge)
Function
Contact type
External delete
Programmable:
Deleting / ac-
break contact or
knowledging exist-
closing contact
ing errors / alarms
(isolated)
Via these terminals, all error messages can be acknowledged re-
motely. The contact type break contact or closing contact can be set
in the menu.
Basic program
Here select =>
Input water meter
Function
Contact type
closing con-
WM
tact/break con-
Water meter
tact
input
(isolated!)
Programming the water meter inputs
Select in the menu =>
Select the meter constant of the water meter
Confirm the input with "OK"
As required, configure the contact type (break contact/closing con-
tact) of the water meter in the menu =>
function WM
.
Current loop 0/4 - 20 mA
Function
OUT
Current loop
0/4 - 20 mA
Loading the current loop
The maximum load of 500 Ohm must not be exceeded!
In the event of faults and very long lines (around 20 m), the use of
shielded cable is possible.
Input basic programming data
Testing
duration
none
Ext. Ack.
=>
Testing
Action
duration
none
Determining quantity
to trigger analysis
Basic program
=>
Basic program
Connection
Action
Burden max.
programmable:
500 Ohm
0 - 20 mA or
4 - 20 mA
Action
Proceed as for
horn button
.
Water meter
=>
47