Quantity control
Time priority
MODE OF OPERATION
Time controlled
Volume interval
Volume and Time
INTERVAL
Time
Volume
03501
INTERVAL
Time
Volume
00001
NOTE
Select quantity control/time priority
The analysis is performed once the programmed quantity of water
has passed. An analysis is always prioritized once the programmed
interval pause has elapsed.
Select the function
Select in the menu =>
ation
*
Confirm the selection with "OK"
(An " * " appears at the end of the line)
Select in the menu =>
Time
10m
Confirm the selection with "OK"
Select the pause time in minutes with the cursor keys.
(The factory default setting is 10 minutes)
Confirm the input with "OK"
Volume
Select
Confirm the selection with "OK"
Input the corresponding flow volume in liters
Confirm the input with "OK"
Configure interval (interval pause)
With time-controlled analysis trigger, the interval between two anal-
yses is determined by the interval pause (plus purge cycle). The
shortest possible interval pause is 0 minutes. Non-stop analyses are
then performed. The largest interval is 99 minutes.
Select in the menu =>
Time
10m
Select the pause time in minutes with the cursor keys.
(The factory default setting is 10 minutes)
Confirm all entries with "OK"
Duration of the analysis interval
The time for the analysis interval comprises the cumulative
time taken for the "Analysis interval", "Purge" and the dura-
tion of analysis, which depends on the measurement value
(see accompanying diagram).
When you perform additional purging via the AUX relay be-
fore or after the analysis, the duration of the analysis interval
will be extended accordingly.
Basic program
Volume and Time
=>
Basic program
with the cursor keys
Basic program
Input basic programming data
Mode of oper-
=>
Interval
=>
Interval
=>
=>
=>
35