Dräger Alcotest 4000 Instrucciones De Uso página 14

Ocultar thumbs Ver también para Alcotest 4000:
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 21
en
|
Disposal
8
Disposal
This product must not be disposed of as household
waste. This is indicated by the adjacent symbol.
You can return this product to Dräger free of charge. For
information please contact the national marketing orga-
nizations or Dräger.
Batteries must not be disposed of as household waste.
They are therefore marked with the adjacent symbol.
Dispose of batteries at battery collection centres as
specified by the applicable regulations.
9
Technical data
9.1
Technical data
Measuring principle
Electrochemical sensor
Measuring range
Breath alcohol concentra-
tion
Blood alcohol concentra-
tion
A conversion factor of 2100 is used to calculate the blood
alcohol concentration using the breath alcohol concentration
data (factory setting, country-specific conversion factors can
be configured).
Measurement units
mg/L, µg/100 mL, µg%, µg/L, g/210 L, g/L ‰, %,
mg/100 mL, mg%, mg/mL
The device meets the requirements of EN 15964. It can be
configured in accordance with the standard.
Display of measured values if configured in accordance with
EN 15964:
Measured value
0.0 mg/L to 0.029 mg/L
0.03 mg/L to 2.5 mg/L
Ambient conditions
During operation
During storage
Sampling
Breath volume
Breath flow
Breath duration
Measurement accuracy
Standard deviation of repeatability with ethanol standard
14
0.00 to 2.50 mg/L
(Ethanol mass per breath volume
at 34 °C and 1013 hPa)
0.00 to 5.00 ‰
(Ethanol mass per blood volume
or mass at 20 °C and 1013 hPa)
Display
0.0 mg/L
Measured value
-5 °C to +50 °C
(23 °F to 122 °F)
10 to 100 % RH
(non-condensing)
600 to 1300 hPa
-20 °C to +60 °C
(-4 °F to +140 °F)
15 to 75 % RH (10 to 100 % RH
<48 h)
>1.2 L (default setting)
>10 L/min (default setting)
>1.6 s (default setting)
Up to 0.50 mg/L
≥0.50 mg/L
Up to 1.00 ‰
≥1.00 ‰
Sensitivity drift
Wait time for measurement readiness following a previ-
ous measurement of:
0.00 mg/L: 2 s
0.25 mg/L: 20 s
0.50 mg/L: 40 s
1.00 mg/L: 80 s
>1.50 mg/L: 120 s
Adjustment
Recommendation: every 12 months
Dimensions
50 mm × 133 mm × 29 mm (W × H × D)
Weight
approx. 130 g
Power supply
1 × 3 V CR123A
The battery is designed for up to 1500 measurements
(depending on use and ambient conditions)
Protection class
IP 52
Interface
Micro-USB port, for service personnel only
CE marking
Electromagnetic compatibility
Configuration possible, see nameplate
Instructions for use
0.008 mg/L or
1.7 % of the measured value
0.017 ‰ or
1.7 % of the measured value
depending on which value is
higher
Typically 0.4 % of the measured
value per month
0.00 ‰: 2 s
0.50 ‰: 20 s
1.00 ‰: 40 s
2.00 ‰: 80 s
>3.00 ‰: 120 s
|
Dräger Alcotest 4000
loading