Dräger Alcotest 6000 Instrucciones De Uso página 21

Ocultar thumbs Ver también para Alcotest 6000:
Idiomas disponibles

Idiomas disponibles

Respiratory volume
Breath flow
Duration of breath
Measurement accuracy
Standard deviation of the repeatability with ethyl alcohol standard
<0.50 mg/L
≥0.50 mg/L
<0.100 % BAC
≥0.100 % BAC
Wait time for measurement readiness following a previous mea-
surement with:
0.00 mg/L: 2 s
0.25 mg/L: 20 s
0.50 mg/L: 40 s
1.00 mg/L: 80 s
>1.50 mg/L: 120 s
Test for accuracy
Recommendation: every 12 months (local deviations are possible)
Dimensions
50/60 mm x 141 mm x 31 mm (W x H x D)
1.92/2.36 in x 5.55 in x 1.22 in (W x H x D)
Weight
approx. 5.3 oz (150 g)
Power supply
1 x 3 V CR123A
The battery is designed for up to 1500 measurements (depending on
use and the ambient conditions)
Interface
Micro-USB, 2.0 fullspeed, not for charging the device, for service only
Protection class
IP 54
Interface
Dräger Alcotest 6000
>1.2 L (default)
>10 L/min (default)
>1.6 s (default)
0.008 mg/L or
1.7 % of the measured value
0.0017 % BAC or
1.7 % of the measured value.
Whichever value is higher
0.00 %: 2 s
0.050 %: 20 s
0.100 %: 40 s
0.200 %: 80 s
>0.300 %: 120 s
|
Disposal