Technical Specifications - Beurer BM 85 Manual De Instrucciones

Tensiómetro
Ocultar thumbs Ver también para BM 85:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 55
Device charged (25% - 50%)
< 25%
You can take a maximum of 10 measurements
(flashes)

12. Technical specifications

Model no.
BM 85
Measurement
Oscillometric, non-invasive blood pres-
method
sure measurement on the upper arm
Measurement range
Cuff pressure 0 – 300 mmHg,
systolic 60 – 260 mmHg,
diastolic 40 – 199 mmHg,
pulse 40 –180 beats /minute
Display accuracy
Systolic ± 3 mmHg,
diastolic ± 3 mmHg,
pulse ± 5 % of the value shown
Measurement inac-
Max. permissible standard deviation
curacy
according to clinical testing:
systolic 8 mmHg /
diastolic 8 mmHg
Memory
2 x 60 memory spaces
Dimensions
L 180 mm x W 99 mm x H 40 mm
Weight
Approximately 456 g (with cuff)
Cuff size
22 to 42 cm
Permissible operating
conditions
Permissible storage
conditions
Power supply
Battery life
Accessories
Classification
Data transfer
Technical information is subject to change without notification
to allow for updates.
The serial number is located on the device or in the battery
compartment.
35
+ 10 °C to + 40 °C, ≤ 90 % relative humi-
dity (non-condensing)
-20 °C to + 55 °C, ≤ 90 % relative humi-
dity, 800 –1050 hPa ambient pressure
DC 5 V
600mA lithium ion battery
3.7 V/400mAh
For approx. 50 measurements, depen-
ding on levels of blood pressure and
pump pressure
Cuff, cuff holder, instructions for use,
USB mains part, USB cable, storage
pouch
Internal supply, IPX0, no AP or APG,
continuous operation, application part
type BF
2402 MHz – 2480 MHz frequency band
Transmission power max. 2.8 dBM
The blood pressure monitor uses Blue­
tooth
low energy technology
®
Compatible with Bluetooth
phones/tablets
List of supported smartphones/tablets
4.0 smart-
®
Tabla de contenido
loading

Tabla de contenido