X-ray emitter
Nominal voltage, high-voltage generator
Nominal current, high voltage generator
Cooling, high-voltage generator
Additional filtering at 50 kV
Integrated filtering at 50 kV
Total filtering at 50 kV
X-ray tube model
Focal spot size as per IEC 60336 X-ray
tube
Anode angle
Anode heat capacity
Pulse/break ratio
Duration of radiation exposure
Maximum current-time product per hour
4.1
X-ray tube performance data
– Maximum deviation of the peak voltage from the displayed value ± 10%
– Maximum deviation of the tube current from the displayed value ± 20%
– Maximum deviation of the exposure time from the displayed value ± 10%
– The device complies with the standards IEC 60601-1, IEC 60601-1-3 and IEC 60601-2-63.
– The lowest possible load factor is obtained with a combination of the settings 60 kV and 4 mA.
2210200496L41 1911V006
kV
4 - 16 (± 10%, max. 75 kV 16 mA,
mA
Automatic monitoring
2.0 + 3.0 (added automatically for CBCT)
mm Al
0.08 + 0.12 (added automatically for
in Al
mm Al
in Al
2.8 + 3.0 (added automatically for CBCT)
mm Al
0.11 + 0.12 (added automatically for
in Al
mm
in
°
kJ
s
mAs
960 (at 75 kV/16 mA)
Product description
50 - 99 (±10 %)
max. 99 kV 10 mA)
Shut-off at ≥ 60°C
CBCT)
0.8
0.03
CBCT)
Toshiba D-052SB
0.5
0.02
5
35
1:60 or more
0.5 - 20
EN-
US
11