Usage
EN-
Comply with the exposure times of the cleaning
US
agent.
Intermediate rinsing
After the action time prescribed by the manufac-
turer has expired:
Rinse all components with water for at least 1
minute (temperature < 95°F).
Disinfecting
Wipe off all surfaces for at least one minute
with a disinfection wipe.
Drying
If necessary, re-dry at a clean location using a
hygienic, lint-free cloth.
Blow dry the components with compressed air
in a clean location.
9.6
Automatic cleaning, inter-
mediate rinsing, disinfection,
final rinse, drying
Selection of the washer-disinfector
Automatic cleaning and disinfection requires a
washer-disinfector with the following properties
and validated processes:
– Corresponds to and tested in accordance with
EN ISO 15883
– Certified program for thermal disinfection (A
value ³ 3000 or at least 5 minutes at 382°F)
Program is suitable for the components and
includes sufficient rinsing cycles.
For more information: "9.3 General information".
Selection of the cleaning agents for automatic
cleaning
The following properties are required:
– Material compatibility with the product
– Compliance with the washer-disinfector manu-
facturer's specifications
For further information, see "9.3 General informa-
tion".
Cleaning and disinfection
Place all components in the washer-disinfector
(follow the manufacturer's instructions).
Make sure there are no hidden areas that are
missed by the rinsing process.
Secure the components with a suitable fixture
of the washer-disinfector.
44
9.7
Check for function
After the end of the cleaning and disinfection
cycle, check the components for any residual
soiling and residual moisture. If necessary,
repeat the cycle.
If necessary, replace any damaged parts.
The components should be packaged as soon
as possible after drying and checking.
9.8
Steam sterilization
Packing
For packing of the components, only use trans-
parent paper-film sterilization packaging that is
approved for use in steam sterilization according
to the manufacturer's instructions. This includes:
– Temperature resistance up to 281°F
– Standards DIN EN ISO 11607-1/2
– The applicable sections of the standard series
DIN EN 868
The sterilization packaging must be sufficiently
large. Once it is loaded, the sterilization packag-
ing may not be under any strain.
Steam sterilization
WARNING
Incorrect sterilization reduces effec-
tiveness and can damage the product
0
Only steam sterilization is permitted.
Comply with the specified process
parameters.
Comply with the manufacturer's
instructions regarding use of the steam
sterilizer.
Do not use any other methods.
Steam sterilizer requirements:
– Complies with EN 13060 or EN 285 and/or
ANSI AAMI ST79
– Suitable programs for the products listed
(e. g. with hollow bodies: fractionated vacuum
procedure including three vacuum steps)
– Sufficient drying of the product
– Validated processes in accordance with ISO
17665 (valid IQ/OQ and product-specific per-
formance appraisal (PQ))
2210200496L41 1911V006