10 Maintenance
US
10.1
Recommended maintenance schedule
Only qualified personnel or personnel trained by Air Techniques is allowed to service the unit.
DANGER
Risk of electric shock due to live parts
Install an all-pole disconnect switch with a contact opening width of at least 3 mm in the elec-
trical connection to the mains power supply. It must be possible to secure the disconnect
switch so that it cannot be inadvertently switched back on again.
WARNING
Risk of infection from contaminated products
Risk of cross contamination
Reprocess the product correctly and promptly before its first use and after every subsequent
use.
NOTICE
Damage to the X-ray tube due to overheating
Observe the cooling curves of the X-ray tube when working with the service tool.
Inspection inter-
val
Every 3 years
46
Inspection work
❯ Functional test of the display. Are all symbols displayed?
❯ Functional test of the exposure button.
❯ Do the various status LEDs light up?
❯ Check that the mechanism of the head supports works correctly. Are the head
supports easy to detach and put back on?
❯ Functional test of the EMERGENCY OFF button. Is the EMERGENCY OFF but-
ton easy to operate mechanically, and does it light up when pressed?
❯ Light barrier test for all light barriers installed in the unit.
❯ Visually check the positioning beams. Check the operation of the levers for
adjusting the beam localizers.
❯ Check the X-ray images for artifacts. If necessary, adjust the collimator and/or
calibrate the sensor.
❯ Check the firmware and software versions.
❯ Perform a comparative dose measurement based on the requirements from
the acceptance test.
❯ Recurring tests and tests after repairs to medical electrical equipment –
IEC 62353 (VDE 0751-1).