8.2 X-ray unit settings
Intraoral X-ray units
The indicated standard values are a recommendation, the dentist determines the individual setting of
the exposure values
A setting of 60 kV is preferred provided it can be set on the X-ray unit
The standard exposure values for F-speed film (e g Kodak Insight) can be used
The following table shows the standard values for the exposure time and the dose area product of a
PSP for an adult patient
Incisor
Premolar
Molar
Bite wing
Incisor
Premolar
Molar
Bite wing
The following table shows the standard values for the exposure time and the dose area product of a
PSP for a pediatric patient
Children are more sensitive to radiation than adults Keep the exposure parameters as low as
possible taking into consideration the image quality Refer also to the the FDA information on
pediatric X-ray imaging
(refer to the FDA website on this topic, see https://tinyurlcom/FDAPediatric)
Incisor
Premolar
Molar
Bite wing
2142100008L41 1708V002
OBJ_DOKU-0000002448-001-mehrere.indd 21
without X-ray field
limitation
60 kV
mGycm
2
008 s
146
012 s
219
017 s
311
018 s
329
without X-ray field
limitation
70 kV
mGycm
2
013 s
118
018 s
164
025 s
228
027 s
246
without X-ray field
limitation
60 kV
mGycm
2
005 s
91
007 s
128
011 s
201
011 s
201
DC emitter, 7 mA
Tube length 20 cm
X-ray field limitation
0.8 x 1.2 in
2
(20 x 30 mm
)
2
60 kV
mGycm
2
008 s
31
012 s
46
017 s
66
018 s
70
DC emitter, 6 mA
Tube length 30 cm
X-ray field limitation
0.8 x 1.2 in
2
(20 x 30 mm
)
2
70 kV
mGycm
2
013 s
25
018 s
34
025 s
48
027 s
52
DC emitter, 7 mA
Tube length 20 cm
X-ray field limitation
0.8 x 1.2 in
2
(20 x 30 mm
)
2
60 kV
mGycm
2
005 s
19
007 s
27
011 s
42
011 s
42
Installation
X-ray field limitation
1.2 x 1.6 in
2
(30 x 40 mm
)
2
60 kV
mGycm
2
008 s
62
012 s
93
017 s
132
018 s
14
X-ray field limitation
1.2 x 1.6 in
2
(30 x 40 mm
)
2
70 kV
mGycm
2
013 s
50
018 s
69
025 s
96
027 s
104
X-ray field limitation
1.2 x 1.6 in
2
(30 x 40 mm
)
2
60 kV
mGycm
2
005 s
38
007 s
54
011 s
85
011 s
85
21