Air Techniques ProVecta 3D Prime Instrucciones De Uso página 18

Ocultar thumbs Ver también para ProVecta 3D Prime:
Idiomas disponibles

Idiomas disponibles

Product description
EN-
Electromagnetic compatibility (EMC)
US
Interference emission measurements
High-frequency emissions in accordance with CISPR 11
Interference voltage at the power supply connection
CISPR 11:2009+A1:2010
Electromagnetic interference radiation
CISPR 11:2009+A1:2010
Emission of harmonics
IEC 61000-3-2:2005+A1:2008+A2:2009
Voltage changes, voltage fluctuations, and emission of
flicker
IEC 61000-3-3:2013
Electromagnetic compatibility (EMC)
Interference immunity measurements on cover
Immunity to interference by discharge of static electricity
IEC 61000-4-2:2008
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air
Immunity to interference by high-frequency electromag-
netic fields
IEC 61000-4-3:2006+A1:2007+A2:2010
3 V/m
80 MHz - 2.7 GHz
80 % AM at 1 kHz
Immunity to interference by near fields of wireless HF
communication devices
IEC 61000-4-3:2006+A1:2007+A2:2010
See Table on immunity to interference table by near fields
of wireless HF communication devices.
Immunity to interference levels by near fields of wireless HF communication devices
Radio service
TETRA 400
GMRS 460
FRS 460
LTE band 13, 17
GSM 800/900
TETRA 800
iDEN 820
CDMA 850
LTE band 5
14
Group 1
Class A
Class A
Class A
Conforms
Conforms
Conforms
Conforms
Frequency band
MHz
380 - 390
430 - 470
704 - 787
800 - 960
2210200496L41 1911V006
Test level
V/m
27
28
9
28
loading