Product description
X-ray emitter
Model
Rated power
Type of high-voltage generator
Nominal voltage, high-voltage generator
Nominal current, high voltage generator
EN
US
Cooling, high-voltage generator
Additional filtration at 50 kV
Integrated filtration at 50 kV
Total filtration at 50 kV
X-ray tube model
Focal spot size as per IEC 60336 X-ray
tube
Anode angle
Anode heat capacity
Pulse/break ratio
Duration of radiation exposure
Maximum current-time product per hour

4.1 X-ray tube performance data

– Maximum deviation of the voltage peak from the displayed value ± 10%
– Maximum deviation of the tube current from the displayed value ± 20%
– Maximum deviation of the exposure time from the displayed value ± 10%
– The device complies with the standards IEC 60601-1, IEC 60601-1-3 and IEC 60601-2-63�
– The lowest possible stress factor is obtained with a combination of the settings 60 kV and 4 mA�
10
OBJ_DOKU-0000003575-mehrere-005.indd 10
kW
kV
mA
mm Al
in Al
mm Al
in Al
mm Al
in Al
mm
in
°
kJ
s
mAs
Maximum Rating Charts
DC (Center Grounded)
Constant potential high-voltage generator
Nominal Focus Spot Value: 0.5
25
70kV
80kV
20
90kV
100kV
15
10
5
1
2
3
EXPOSURE TIME [s]
DG-07C11T2 (H)
1�6 (for 1 s)
Inverter
50 - 99 (±10 %)
4 - 16 (± 10%, max� 75 kV 16 mA,
max� 99 kV 10 mA)
Automatic monitoring
Shut-off at ≥ 60°C
2�0 + 3�0 (autom� added for CBCT)
0�08 + 0�12 (autom� added for CBCT)
0�03
2�8 + 3�0 (autom� added for CBCT)
0�11 + 0�12 (autom� added for CBCT)
Toshiba D-052SB
0�02
1:60 or more
0�5 - 20
960 (at 75 kV/16 mA)
60kV
50kV
5
7
10
20
2210200496L41 1904V004
0�8
0�5
5
35
17.04.2019 14:28:52