The reprocessing procedure was validated as fol-
lows:
– Pre-cleaning:
– Monarch surface disinfection wipes (Air
Techniques)
– Cleaning brush
– Manual cleaning:
– Monarch enzymatic cleaner 2.0 % (v/v) (Air
Techniques)
– Manual disinfection:
– Monarch surface disinfection wipes (Air
Techniques)
– Automated cleaning and disinfection was per-
formed in accordance with EN ISO 15883 with
tested efficacy:
– Washer-disinfector PG 8536 CD (Miele,
Gütersloh, Germany)
– Cleaning agent: Neodisher MediClean Forte
– Programs: Cleaning without neutralization
and THERMAL DES
– Steam sterilization:
– Steam sterilizer Systec DX-45 (Systec
GmbH, Linden, Germany)
9.3
General information
Comply with all national directives, standards
and specifications for the cleaning, disinfection
and sterilization of medical devices as well as
the specific specifications for dental practices
and clinics.
Comply with the specifications (in "9.5 Manual
cleaning, intermediate rinsing, disinfection, dry-
ing" and "9.6 Automatic cleaning, intermediate
rinsing, disinfection, final rinse, drying") when
you select the cleaning agents and disinfec-
tants to be used.
Comply with the concentrations, temperatures,
residence times and post-rinsing specifications
issued by the manufacturer of the cleaning
agent and disinfectant.
Only use cleaning agents that are non-fixing
and aldehyde-free and display material com-
patibility with the product.
Only use disinfectants that are aldehyde-free
and display material compatibility with the
product.
Do not use any rinse aid (danger of toxic resi-
due on the components).
Only use freshly-produced solutions.
2210200496L41 1911V006
Only use distilled or de-ionized water with a low
bacterial count (at least drinking water quality)
that is free from facultatively pathogenic micro-
organisms (e.g. Legionella bacteria).
Use clean, dry, oil- and particle-free com-
pressed air.
Do not exceed temperatures of 281 °F.
Subject all devices used (e.g. ultrasonic bath,
cleaning and disinfection device (washer-disin-
fector), sealing device, steam sterilizer) to regu-
lar maintenance and inspections.
9.4
Preparation at the operating
location
Wear hand protection.
Wear eye protection.
Use a mask.
Wear protective clothing.
WARNING
Risk of infection from contaminated
products
Risk of cross contamination
Reprocess the product correctly and
promptly before its first use and after
every subsequent use.
Protect the unit from contamination when
transporting it from the treatment chair to the
reprocessing location.
Brush off all surfaces below the water surface
with a soft hygienic brush until they are clean to
the eye.
Wipe off all surfaces for at least one minute
with a disinfection wipe.
9.5
Manual cleaning, intermedi-
ate rinsing, disinfection, dry-
ing
Cleaning
Place individual parts in a cleaning agent bath
making sure that all parts are covered.
Usage
EN-
US
43