EN-
8
Cleaning and disinfection
US
NOTICE
The use of unsuitable agents and
methods can damage the unit and
accessories
Only use the disinfection and cleaning
❯
agents specified or approved by Air
Techniques and the EPA.
Comply with the operating instructions
❯
of the disinfectants and cleaning
agents.
Wear hand protection.
Prior to working on the unit or in case of
danger, disconnect it from the mains.
The following cleansers and disinfection agents
have been tested for the compatibility of materi-
als:
– Birex wipes
– Discide Ultra Towelettes
– Opti Cide 3 surface wipes
– Optim 33TB wipes
– Maxiwipe germicidal cloth
– Cavi Wipes
– Cavi Wipes 1
– PDI Sani-Cloth
– Clorox
– Monarch
8.1
Surface of the unit
NOTICE
Damage to the touch screen caused
by cleaning it with disinfectant
Only clean the touch screen with a soft
❯
cloth and a commercially available
cleaning agent.
The surface of the unit must be cleaned and dis-
infected if it is contaminated or soiled.
NOTICE
Liquid can cause damage to the unit
Do not spray the unit with cleaning
❯
agents or disinfectants.
Make sure that liquid penetrates into
❯
the unit.
40
Remove any soiling with a soft, damp, lint-free
❯
cloth.
Disinfect the surfaces with a disinfection wipe.
❯
Alternatively, use a quick-acting surface disin-
fectant on a soft, lint-free cloth. Comply with
the operating instructions of the disinfectant.
8.2
Positioning aids
The positioning aids must be cleaned and disin-
fected if they are contaminated or soiled.
Head supports with cushion
Pull the head supports off the device.
❯
Remove the cushions from the head supports.
❯
Remove the cushion holder.
❯
Remove any soiling with a soft, damp, lint-free
❯
cloth.
2210200496L41 1911V006