Faults, cause and remedy
11
Faults, cause and remedy
Display
Flow detected during Warm-Up
Blow stronger ! / Repeat Test
Blow softer ! / Repeat Test
Blow longer ! / Repeat Test
Suck after Blow / Repeat Test
Suck longer ! / Repeat Test
Check mouthpiece / Repeat Test
Please retry / Repeat Test
Test Failed
Low Temperature / Warm Up Handset
High Temperature / Cool down Handset Temperature of handset too high.
Service required in x days
Service lockout
28
Cause
Breath sample has been blown too early.
Breath sample has been blown too soft.
Breath sample has been blown too strong. Repeat test blowing softer.
Breath sample has been blown too short. Repeat test blowing longer.
Breath sample is not delivered correctly
by blowing and sucking.
Breath sample has been sucked too short Repeat test sucking longer.
Mouthpiece is not correctly inserted into
handset.
During delivery of the breath sample a
fault has been detected.
The measured breath-alcohol
concentration is above the set limit value.
Temperature of the handset too low.
Service interval will expire in x days.
Service interval expired.
Remedy
Deliver breath sample when requested by the
instrument.
Repeat test blowing stronger.
Repeat test with blowing and immediately afterwards
sucking.
Insert mouthpiece correctly.
Deliver breath sample again when requested by the
instrument.
Repeat test after at least 15 minutes.
If message reappears, the breath alcohol concentration
is above the limit value.
Detach handset and take it to a place with a moderate
temperature. Connect measuring instrument again and
make measurement.
Detach handset and take it to a place with a moderate
temperature. Connect measuring instrument again and
make measurement.
Call an authorized Dräger Interlock Service Center in
due time.
Call an authorized Dräger Interlock Service Center.