7.4 ENDOSCOPE INSTALLATION
1. Straighten the Balloon Catheter to enable proper seating of the Endoscope.
2. Pull out the Endoscope fitting on the Balloon Catheter handle and lock it into the extended position by rotating
it one half turn.
3. Remove the cap on the Endoscope fitting, and dispose of cap.
4. Insert the Endoscope through the Endoscope fitting and into the Balloon Catheter. Securely connect the luer
on the Endoscope to the Endoscope fitting.
5. Rotate the Endoscope fitting one half turn to unlock it from the extended position.
6. Pass the proximal end of the Endoscope to the NST for connection to the camera. An image will appear on
the Console display.
7. On the Balloon Remote press the '+' side of the rocker switch to inflate the balloon to a diameter just beyond
wrinkle free balloon surface.
8. Release the rocker switch to evacuate any air in the balloon.
7.5 CONNECT LESION GENERATOR AND MOTOR CABLE
1. The GST should pass the Lesion Generator's optical connector and motor cable to the NST.
2. The NST will un-cap the Lesion Generator's fiber connector. The NST should take care not to allow the optical
connector face to come in contact with anything.
3. The NST should disconnect the dust cap and immediately insert the Lesion Generator connector into the console
output port taking care not to allow the face of the connector to touch the output port. Turn the knurled nut clockwise
until tight.
Caution: Do not contaminate the proximal face of the Lesion Generator connector. Be sure to insert
the Lesion Generator by carefully holding the connector at an angle, tipping the connector into the output
port and inserting the connector into the output port.
4. The NST should plug the motor cable into the motor cable connector on the Console. An audible click will be
heard when connection is complete.
5. Enter the Lesion Generator Serial Number into the Console.
Note: If it is necessary to disconnect the Lesion Generator, it may be reconnected within 24
hours. After 24 hours the Catheter serial number is invalid.
7.6 BALLOON CATHETER PURGING
1. With the pump running, purge all remaining air from the Balloon Catheter as follows: Press the Balloon Remote's
'+' side of the rocker switch to inflate the balloon. The physician or GST should hold the Balloon distal end down and
stop depressing '+' to allow any air to flow out of the Balloon. Allow all air to flow out of the return line. On the
Balloon Remote press the '-' side of the rocker switch to completely deflate the balloon.
2. With the balloon deflated, add Balloon Fill Media to the burette. Fill the burette up to the solid line of the
burette.
HeartLight X3 Catheter Instructions for Use
06-5511 REV A
Page 9
ECN-4541