Philips V60 Guía De Uso página 10

Interfaz de comunicación para vrpt/snda
Ocultar thumbs Ver también para V60:
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 62
Field
Description
11
PEEP or EPAP setting
12 to 21
Unused
22
Pressure support setting
23 to 29
Unused
30
Measured total
respiratory rate
31
Measured tidal volume
32
Measured total minute
volume
33
Unused
34
Measured peak
inhalation pressure
35 to 37
Unused
38
High inhalation pressure
alarm setting
Table 1-2: SNDA record format (continued)
Example
Resolution
0.0◆◆◆
0.1
◆◆◆◆◆◆
N/A
0◆◆◆◆◆
1
◆◆◆◆◆◆
N/A
0.0◆◆◆
0.1 for 1.0
to 9.9
1 for 10 to
90
0.00◆◆
0.01
0.00◆◆
0.01 for
0.00 to
9.99
0.1 for 10.0
to 99.9
◆◆◆◆◆◆
N/A
50.0◆◆
0.1
N/A
◆◆◆◆◆◆
20◆◆◆◆
1
Range
Units
Comments
4.0 to 25.0
cmH
O
CPAP or EPAP
2
setting.
"◆◆◆◆◆◆" in HFT.
In CPAP mode:
EPAP = IPAP = CPAP
N/A
N/A
Always output as
"◆◆◆◆◆◆"
0 to 36
cmH
O
IPAP - EPAP in S/T
2
and PCV modes.
0 in CPAP mode.
"◆◆◆◆◆◆" in
AVAPS and PPV
modes, and HFT.
N/A
N/A
Always output as
"◆◆◆◆◆◆"
0.0 to 9.9
BPM
"◆◆◆◆◆◆" in
standby mode or
during exhalation
10 to 90
port leak test.
"◆◆◆◆◆◆" in HFT
0.00 to 3.50
L
"◆◆◆◆◆◆" in
standby mode, during
exhalation port leak
test, or for values out
of range.
"◆◆◆◆◆◆" in HFT
0.00 to 9.99
L
For values out of
range set output to
99.9◆◆
10.0 to 99.9
"◆◆◆◆◆◆" in
standby mode or
during exhalation
port leak test.
"◆◆◆◆◆◆" in HFT
N/A
N/A
Always output as
"◆◆◆◆◆◆"
0.0 to 50.0
cmH
O
"◆◆◆◆◆◆" in
2
standby mode or
during exhalation
port leak test.
"◆◆◆◆◆◆" in HFT
N/A
N/A
Always output as
"◆◆◆◆◆◆"
5 to 50
cmH
O
"◆◆◆◆◆◆" in HFT
2
9