Relay output

RS485 connection:
1
2
3
1
1
2
XEXT1
2
3
3
CAUTION
!
Ensure that the cable strands are not crossed!
1
1 Plug (male):
Housing: Souriau UTS6JC147P (male)
Contacts: Souriau RM20M12K (male)
2 Plug (female):
Housing: Souriau UTS6JC147S (female)
Contacts: Souriau RC20M12K (female)
3 Cable type: Belden 3107A, 2 x 2 AWG 22
Cable length: max. 25 m between 2 devices
50
1
XEXT2
2
3
1
2
3
3
max. 25 m
6
Relay output
The relay output is intended for connection to an isolation
amplifier with intrinsically safe output.
1
The behaviour of the relay output can be configured between
2
the NC and NO setting using the Dräger CC-Vision PC
3
software.
Factory setting of the relay output: NC
02333112.eps
State of X-zone: State of relay
Off
Operation without
alarms
Alarms
2
WARNING
!
The behaviour of the relay output must be taken into
account when carrying out the safety design of the
sequential switching.
It is essential that the parameters of the relay outputs
03133112.eps
are observed to prevent the circuit from jeopardising
the intrinsic safety of the device.
Only intrinsically safe circuits are permitted for
connection. The cable shielding must only be placed
onto the isolation amplifier output. Connection XEXT2
must be used either for the relay output only or an
RS485 connection. It is not permissible to use the
connector for both purposes at the same time!
The cables used for the alarm contact must conform to
the
requirements
IEC 60079-11 and IEC 60079-14. The following cable
conforms to these standards: Belden 3107A.
NOTICE
i
i
The person responsible for the use of the X-zone 5000
must prepare a system document to verify the intrinsic
safety level.
NO
5
NC
4
State of relay
output: config. NC
output: config. NO
4 connected to 6
5 connected to 6
4 connected to 6
5 connected to 6
5 connected to 6
4 connected to 6
of
standards
Dräger X-zone 5000
CO
6
IEC
60079-0,