Interval
Daily or as required
Weekly or as required
Annually or as required
According to the applicable national regula-
tions
10. Rectification of faults and malfunctions
10.1.
FAULTS TABLE
Fault
No camera image, or the camera
image is black.
Device cannot be calibrated.
No transillumination.
Counter shows unrealistic coordin-
ates.
Neither quick adjustment nor fine
adjustment functioning.
No printing.
Panel PC failing to start up.
Maintenance work
To avoid corrosion, depending on the work-
ing environment, daily grease the tool
mounting after use, or rub a lubricant such
as WD-40 into it.
Wipe down the housing with a slightly
damp cloth.
The optical components such as the lens
and camera should be kept free of oil, dust
and fingerprints. Clean them with a non-
linting cloth to which a little alcohol has
been applied.
Clean the screen with a commercially-avail-
able glass cleaner and a soft cloth.
1. Move the guided units to their respective
end positions.
2. Inspect the guides for the tower and anti-
rotation arm and grease them as necessary.
» Use multi-purpose grease. Recommend-
ation: LGEP 2/1 from SKF.
Check the calibration and check the guides
for play.
Check the electrical components (including the
power supply unit) for conformity to national
regulations.
Possible cause
Camera software has crashed.
Cable open circuit
Incorrect lighting setting.
No calibration master gauge in the
live image
LED defective
Cable open circuit
Switched off in the PCTC software Restart the panel PC.
Movement to reference points
failed
Communications problem with the
reading head
Adapter incorrectly created or in-
correct adapter selected
Compressed air not connected
Drive nut defective
Pinion displaced
Printer not switched on.
Not connected to the PC.
Not selected in the measure-
ment software.
Not yet installed.
No power supply.
Loose plug at the PC.
www.hoffmann-group.com
Performed by
Trained person
Trained person
Trained specialist for mechanical work
Action
Restart the panel PC.
Check the plug connections.
Check the lighting setting.
1. Move the adapter/calibration
mandrel into the measurement
window.
2. Use the dynamic focus display
[} Page 22] to move the ad-
apter/calibration mandrel into
focus.
Replace the LED.
Check the plug connections.
Repeat the movement to refer-
ence points [} Page 20].
Check or clean the magnetic strip
installed in the tower and the read-
ing head on the guide rails.
Check adapter data, or recreate or
recalibrate it.
Check the compressed air connec-
tions and the service unit.
Check the drive nuts.
Check the position of the pinion.
Check all plug connections.
Check whether the printer is se-
lected in the software.
Install the printer.
Check the plug connections (when
power is present, the green LED on
the power supply unit lights up).
Performed by
Trained person
Trained person
Trained person
Trained person
Trained specialist for mechanical
work
Trained person
Trained person
en