Warning: The BIOFLEX
Personal Therapy System should not be
®
used adjacent to or stacked with other equipment and that if adjacent
or stacked use is necessary, the equipment should be observed to verify
normal operation in the configuration in which it will be used.
Table 2 – Example Emission Class and Group Compliance
The BIOFLEX
Personal Therapy System is intended for use in the
®
electromagnetic environment specified below. The customer or the
user of the BIOFLEX
Personal Therapy System should assure that it is
®
used in such an environment.
Emissions Test
Compliance
RF Emissions
Group 1
CISPR 11
RF Emissions
Class B
CISPR 11
Harmonic
Emissions
Class A
IEC 61000-3-2
Voltage
Fluctuations/
Flicker
Complies
Emissions IEC
61000-3-3
38
MN 421.012 Rev 3
Electromagnetic
Environment – Guidance
The BIOFLEX
Personal Therapy
®
System uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not
likely to cause any interference in
nearby electronic equipment.
The BIOFLEX
Personal Therapy
®
System is suitable for use in all
establishments including domestic
establishments and those directly
connected to the public low-
voltage power supply network
that supplies buildings used for
domestic purposes
Table 3 – Immunity Test Level Compliance
The BIOFLEX
Personal Therapy System is intended for use in the electromagnetic
®
environment specified below. The customer or the user of the BIOFLEX
System should assure that it is used in such an environment.
IEC 60601 Test
Immunity Test
Level
Electrostatic
Discharge
±8 kV Contact ±15
(ESD) IEC
kV Air
61000-4-2
Electrical Fast
±2 kV for power
Transient/Burst IEC
supply lines
61000-4-4
Surge IEC 61000-
±0.5 kV, ±1 kV line
4-5
to line
0% U
T
(100% dip in UT)
for 0,5 cycle
0% U
Voltage Dips, short
T
(100% dip in U
)
interruptions and
T
for 1 cycle
voltage variations
on power supply
70% U
input lines
T
(30% dip in UT)
for 25/30 cycles
IEC 61000-4-11
0% U
T
(100% dip in UT)
for 5 sec
Power
Frequency
Magnetic Field
30 A/m
(50/60 Hz)
IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
Personal Therapy
®
Electromagnetic
Compliance Level
Environment –
Guidance
Floors should be wood,
concrete or ceramic tile.
±8 kV Contact ±15
If floors are covered with
kV Air
synthetic material, the
relative humidity should
be at least 30%.
Mains power quality
±2 kV for power
should be that of a
supply lines
typical commercial or
hospital environment.
Mains power quality
±0.5 kV, ±1 kV line
should be that of a
to line
typical commercial or
hospital environment.
0% U
Mains power quality
T
(100% dip in UT)
should be that of a
for 0,5 cycle
typical commercial or
hospital environment. If
0% U
the user of the BIOFLEX
T
(100% dip in UT)
Personal Therapy System
for 1 cycle
requires continued
operation during power
70% U
mains interruptions, it
T
(30% dip in UT)
is recommended that
for 25/30 cycles
the BIOFLEX
Personal
®
Therapy System be
0% U
powered from an
T
(100% dip in UT)
uninterruptible power
for 5 sec
supply or a battery.
Power frequency
magnetic fields should
be at levels characteristic
30 A/m
of a typical location in
a typical commercial or
hospital environment.
MN 421.012 Rev 3
®