Making Measurements
% Scale of 4-20 mA or 0-20 mA
The 4-20 mA current loop output from a transmitter is a type of electrical signal
that is used in a series circuit to provide a robust measurement signal that is
proportional to the applied pressure, temperature, or flow in process control. The
signal is a current loop where 4 mA represents the zero percent signal and 20 mA
represents the 100 percent signal.
The % scale for 4-20 mA or 0-20 mA in this multimeter is calculated using its
corresponding DC mA measurement. The multimeter will automatically optimize
the best resolution for the selected measurement. Two ranges are available for the
% scale as shown in
To display the current measurement in % scale:
1 Position your multimeter's rotary switch position to
2 Press
Figure 2-34
The analog bar graph displays the current measurement value. (In the example
above, 8 mA is represented as 25% in the 4-20 mA % scale.)
98
Table
multimeter to measure DC current by following the steps listed in the
Measuring AC or DC Current
until
Probe the test points and read the display.
4-20 mA % Scale display
2-3.
section.
(or
) is shown on the right side of the display.
Keysight U1271A/U1272A User's Guide
. Set up your