Changing the % scale range

This setting is used with % scale current measurements (see
multimeter converts DC current measurements to a percentage scale readout of
0% to 100% based on the selected range in this menu. For example, a 25%
readout represents a DC current of 8 mA on the 4-20 mA % scale, or a DC current
of 5 mA on the 0-20 mA % scale.
Parameter
PErCEn
To change the % scale range:
1 Press
2 Press
Figure 4-11
3 Press
4 Press
5 Press and hold
Keysight U1271A/U1272A User's Guide
Range
4-20 mA, 0-20 mA, or Off
for more than 1 second to enter the multimeter's setup menu.
Setup
or
PErCEn display
or
to change the % scale range. Select
Dual
E xit
Setup
scale readout.
to save your changes or press
until the multimeter restarts to return to normal operation.
Setup
until
is shown on the secondary display.
Multimeter Setup Options
page
98). The
Defaul t setting
4-20 mA
to disable the %
to discard your changes.
4
141