Multimeter Features
3
Figure 3-7
Interval log display
2 Press
for more than 1 second to exit the interval log mode.
The maximum number of readings that can be stored for the interval log is
10000 entries for the U1272A and 200 entries for the U1271A. When all entries
are occupied,
will be shown when
is pressed.
The interval and event log share the same memory buffer (total combined:
10000 entries for U1272A and 200 entries for U1271A). Increased usage of the
interval log entries will lead to the decrease of the maximum entries for the event
log, and vice versa.
See the
Reviewing Previously Recorded Data (View)
section later in this manual to
review or erase the recorded entries.
When the interval log recording session is running, all other keypad operations
NOTE
are disabled; except for
, which, when pressed for more than 1 second, will
stop and exit the recording session. Furthermore, APO (auto power-off) is
disabled during the recording session.
Keysight U1271A/U1272A User's Guide
119