Measuring Conductance
Conductance is the reciprocal of resistance. High values of conductance
correspond to low values of resistance. Conductance is measured in Siemens (S).
The 300 nS range measures conductance in nano-Siemens
(1 nS = 0.000000001 Siemens). Because small conductance values correspond to
extremely high resistance values, the nS range allows you to easily calculate and
determine the resistance of components up to 100 GΩ (0.01 nS resolution).
1 To measure conductance, set the multimeter's rotary switch to
2 Press
High-resistance readings are susceptible to electrical noise. Use averaging to
smooth out most of the noisy readings. Refer to
Minimum Values (MaxMin)"
Keysight U1271A/U1272A User's Guide
set up your multimeter as shown in
until the conductance measurement is selected (nS unit shown).
Probe the test points and read the display.
Figure
2-14.
on page 112.
Making Measurements
Smart
"Capturing Maximum and
2
/
and
73