Nouvag MD11 Instrucciones De Uso página 65

Ocultar thumbs Ver también para MD11:
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 133
MD 11
MD11, Error-messages on Display
Error-messages/Error Code
Basic Initialization/
Woo
Set default value/
Wo1
Memory error/
Eo2
Handling error/
E03
Program SW error/
E04
User Config SW error/
E05
Display error/
E06
Pump error/
E07
Storing factory settings/
User Config & Program
Storing factory settings/
Program
Pedal not connected/
E10
Pedal test mode/
W11
Keyboard test mode/
W12
No motor connected/
E13
Unknown motor/
E16
Pump is open/
E20
Motor or pump test mode/
W21
AS-mode torque reached
Pedal locked/
W26, pedal release
Handpiece XX is faulty/
E29
MD 11, V2.0, Operating Instructions, REF 31665, V22/17
Cause
First Initialization
Parameter reset to default value
System Error
System Error
System Error
System Error
System Error
System Error
Message while default values of parameters
and programs saved to the NOU-dongle.
Message while default values of programs are
stored.
a) Pedal not plugged in.
b) Plug or cable is defective.
Pedal test mode switched on
Keyboard test mode switched on
a) No motor connected
b) Motor, motor cable, motor plug or Control
Unit are is defective
a) Wrong motor plugged in.
b) Right motor connected, but motor,
motor cable, motor plug or Control Unit is
defective.
Motor is not working when pump
compartment is open to prevent of injuries.
Motor or pump test mode is switched on.
When the max. torque set is reached in AS-
mode, this message is displayed.
If pedal is pressed at switch on procedure,
pedal will not work.
The handpiece was exposed to excessive
torque, while calibrating or testing.
Solution
Send Control Unit to Service Center.
Send Control Unit to Service Center.
Send Control Unit to Service Center.
Send Control Unit to Service Center.
Send Control Unit to Service Center.
Send Control Unit to Service Center.
a) Plug in Pedal
b) Send Control Unit and pedal to Service
Center.
Switch off device for 5 seconds, than switch
on again.
Switch off device for 5 seconds, than switch
on again.
a) Plug in motor
b) Send motor and Control Unit to Service
Center.
a) Plug in correct motor
b) Send motor and Control Unit to Service
Center.
Close pump compartment.
Switch off device for 5 seconds, than switch
on again.
Release pedal and restart motor by pressing
pedal again.
Release pedal for one second.
-
Clean handpiece and spray it thoroughly
with Nou-Clean spray.
-
If message still displayed after testing,
send handpiece to Service Center.
EN
29
loading