Nouvag MD11 Instrucciones De Uso página 54

Ocultar thumbs Ver también para MD11:
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 133
Handpiece activation
Handpiece existing/HP 02
Handpiece existing/HP 03
Handpiece existing/HP 04
Handpiece existing/HP 05
Parameter Level 2
To change the values in level 2 the password „9403" must be entered. The password can not be changed.
1. Entering the access code: Press Speed ▼.
2. Selecting the access code by pressing HP to increase, or PR to decrease the value.
3. To confirm the access code press Speed ▼ for one second, until a sound is emitted.
(for fast forwarding or backwarding keep keys pressed)
Handpiece
Max. Speed
Handpiece max speed/HP 01
Handpiece max speed/HP 02
Handpiece max speed/HP 03
Handpiece max speed/HP 04
Handpiece max speed/HP 05
AS-Zone
for handpieces
Handpiece AS-Mode/HP 02
Handpiece AS-Mode/HP 03
Handpiece AS-Mode/HP 04
Handpiece AS-Mode/HP 05
Pump parameters
Pump/Backwards
turn mode variable
Pump/Way backwards
Pump/Speed backwards
Pump/Range 1 increment
Pump/Range 1 end
Pump/Range 2 increment
Pump/Range 2 end
Pump/Range 3 increment
Pump/Range 3 end
18
Name of handpiece
Choice
on display
16 : 1
yes/no
20 : 1
yes/no
32 : 1
yes/no
70 : 1
yes/no
Name of handpiece
Speed range
on the display
1 : 1
300 – 40,000
16 : 1
20 – 2400
20 : 1
15 – 1700
32 : 1
10 – 1000
70 : 1
5 – 600
Handpiece
Selection
on display
... to ... rpm
16 : 1
20 – 45
20 : 1
15 – 45
32 : 1
10 – 45
70 : 1
5 – 45
Range
Factory
default
No/Yes
1 – 100 %
10 – 50 %
1 – 10 %
5 – 50 %
1 – 10 %
10 – 90 %
1 – 10 %
20 – 100 %
100 %
Factory
Definition
default
no
Only one handpiece or contra angle can
be selected.
yes
no
no
Factory
Definition
rpm
default
40,000
Limit the maximum speed of your
handpieces according to your own
2400
experience.
1700
1000
600
Factory
Definition
default
20
Selection range for max. AS speed
20
Selection range for max. AS speed
20
Selection range for max. AS speed
20
Selection range for max. AS speed
Definition
Yes
The pressures in the tube set vary according to the
pump speed. In "variable mode" this is considered, to
prevent of spilling when pump is switched off in
backwards mode.
25 %
Specify how far the pump turns backwards
33 %
Specify how fast the pump has to turn backwards to
prevent of spilling after the switch off of the motor.
5 %
Adjustment steps in section 1.
10 %
Set the range where section 1 is active
5 %
Adjustment steps in section 2
50 %
Set the range where section 2 is active.
10 %
Adjustment steps in section 3
Set the range where section 3 is active.
MD 11, V2.0, Operating Instructions, REF 31665, V22/17
MD 11
loading