Functional check
Prior to MD 11 startup or use of accessory equipment, the user must always ensure that each individual
component is in good working order, free from defects, and is clean, sterile and operational. All inscriptions
on the device and its accessories must be readable and there must be no loose parts in the device. Once the
device is switched on, the most recent settings entered appear on the display.
7.10.1
Electronic motor
Use the "Speed ▲" key to set the speed of the electronic motor to its maximum. With the 1 : 1 Contra Angle
attached, it is 40,000 rpm. With all the other Contra Angles the maximum Speed is lower, according to the
transmission ratio of the Contra Angel. Press the pedal treadplate; the electronic motor starts running and
accelerates to 40,000 rpm. When the treadplate is released, the electronic motor slows down again.
The electronic motor is designed for an intermittent duty operation at maximum speed of "1 min ON/3
min OFF" at 4 cycles. Brake after 15 min. Otherwise the electronic motor can be damaged by
excessive heat generation, and by touching the motors hull it could cause serious burnings.
The motor ventilation slots must be kept clear in order to prevent the motor temperature from
becoming excessive.
7.10.2
Pump
Press the key "PUMP" on the pedal briefly; the peristaltic pump is switched on, which is shown on the
display by the symbol of a drop. Press the pedal treadplate; the peristaltic pump and the electronic motor
start up. Water sprays from the irrigation needle on the contra angle.
7.10.3
Rotational direction of the electronic motor
Press the key "FOR/REV" on the pedal briefly; the rotational direction of the electronic motor changes.
Press the pedal treadplate; the electronic motor rotates to the left and a continuous tone is emitted.
Release the treadplate; the electronic motor ceases to operate and the tone is no longer heard. By pressing
the motor key again, the rotational direction is switched back to right rotation, which is shown on the
display by the symbol of a changing arrow.